Quality Inspection for Infusion And Syringe Pump - ZNB-XAII Infusion Pump – KellyMed
Quality Inspection for Infusion And Syringe Pump - ZNB-XAII Infusion Pump – KellyMed Detail:
FAQ
Q: Is the pump open system?
A: Yes, Universal IV set can be used with our Infusion Pump after calibration.
Q: Does the pump compatible with Micro IV Set (1 ml=60 drops)?
A: Yes, all of our pumps are compatible with IV Set of 15/20/60 dorps.
Q: Is it a peristaltic pumping mechanism?
A: Yes, curvilinear peristaltic.
Q: What is the difference between PURGE and BOLUS function?
A: Purge used to remove air in line before infusion. Bolus can be administered for infusion therapy during infusion. Both purge and bolus rate are programmable.
Specifications
| Model | ZNB-XAII |
| Pumping Mechanism | Curvilinear peristaltic |
| IV Set | Compatible with IV sets of any standard |
| Flow Rate | 1-1500 ml/h (in 0.1 ml/h increments) |
| Purge, Bolus | 100-1500 ml/h (in 0.1 ml/h increments)
Purge when pump stops, bolus when pump starts |
| Accuracy | ±3% |
| *Inbuilt Thermostat | 30-45℃, adjustable |
| VTBI | 1-20000 ml (in 0.1 ml increments) |
| Infusion Mode | ml/h, drop/min, time-based, body weight, nutrition |
| KVO Rate | 0.1-5 ml/h (in 0.1 ml/h increments) |
| Alarms | Occlusion, air-in-line, door open, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, standby |
| Additional Features | Real-time infused volume, automatic power switching, mute key,
purge, bolus, system memory, history log, key locker, drug library, rotary knob, change flow rate without stopping the pump |
| Drug Library | Available |
| Occlusion Sensitivity | High, medium, low |
| History Log | 50000 events |
| Air-in-line Detection | Ultrasonic detector |
| Wireless management | Optional |
| Drop Sensor | Optional |
| Vehicle Power (Ambulance) | 12±1.2 V |
| Power Supply, AC | 110/230 V (optional), 50-60 Hz, 20 VA |
| Battery | 9.6±1.6 V, rechargeable |
| Battery Life | 5 hours at 25 ml/h |
| Working Temperature | 10-30℃ |
| Relative Humidity | 30-75% |
| Atmospheric Pressure | 860-1060 hpa |
| Size | 130*145*228 mm |
| Weight | 2.5 kg |
| Safety Classification | Class Ⅰ, type CF |
Product detail pictures:
Related Product Guide:
Our target is to consolidate and improve the quality and service of existing products, meanwhile constantly develop new products to meet different customers' demands for Quality Inspection for Infusion And Syringe Pump - ZNB-XAII Infusion Pump – KellyMed , The product will supply to all over the world, such as: Afghanistan, Netherlands, Stuttgart, To make every client satisfied with us and achieve win-win success, we will continue to try our best to serve and satisfy you! Sincerely looking forward to cooperating with more overseas customers based on mutual benefits and great future business. Thank you.