Manufacturer for Intermittent Pneumatic Compression Stockings - KL-605T TCI Pump – KellyMed
Manufacturer for Intermittent Pneumatic Compression Stockings - KL-605T TCI Pump – KellyMed Detail:
Specifications
| Model | KL-605T |
| Syringe Size | 5, 10, 20, 30, 50/60 ml |
| Applicable Syringe | Compatible with syringe of any standard |
| VTBI | 1-1000 ml (in 0.1, 1, 10 ml increments) |
| Flow Rate | Syringe 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)
Syringe 10 ml: 0.1-300 ml/h Syringe 20 ml: 0.1-600 ml/h Syringe 30 ml: 0.1-800 ml/h Syringe 50/60 ml: 0.1-1200 ml/h |
| Bolus Rate | 5 ml: 0.1-100 ml/h (in 0.01, 0.1, 1, 10 ml/h increments)
10 ml: 0.1-300 ml/h 20 ml: 0.1-600 ml/h 30 ml: 0.1-800 ml/h 50/60 ml: 0.1-1200 ml/h |
| Anti-Bolus | Automatic |
| Accuracy | ±2% (mechanical accuracy≤1%) |
| Infusion Mode | 1. Easy mode
2. Flow rate 3. Time-based 4. Body weight 5. Plasma TCI 6. Effect TCI |
| KVO Rate | 0.1-1 ml/h (in 0.01 ml/h increments) |
| Alarms | Occlusion, near empty, end program, low battery, end battery,
AC power off, motor malfunction, system malfunction, standby, pressure sensor error, syringe installation error, syringe drop off |
| Additional Features | Real-time infused volume, automatic power switching,
automatic syringe identification, mute key, purge, bolus, anti-bolus, system memory, history log |
| Drug Library | Available |
| Occlusion Sensitivity | High, medium, low |
| History Log | 50000 events |
| Power Supply, AC | 110-230 V, 50/60 Hz, 20 VA |
| Battery | 14.8 V, rechargeable |
| Battery Life | 8 hours at 5 ml/h |
| Working Temperature | 5-40℃ |
| Relative Humidity | 20-90% |
| Atmospheric Pressure | 700-1060 hpa |
| Size | 245*120*115 mm |
| Weight | 2.5 kg |
| Safety Classification | Class Ⅱ, type BF |
Product detail pictures:
Related Product Guide:
With state-of-the-art technologies and facilities, strict good quality regulate, reasonable cost, exceptional assistance and close co-operation with prospects, we're devoted to supplying the top benefit for our customers for Manufacturer for Intermittent Pneumatic Compression Stockings - KL-605T TCI Pump – KellyMed , The product will supply to all over the world, such as: Karachi, Bangladesh, Seattle, If any item be of interest to you, please let us know. We will try our best to satisfy your requirements with high quality products, the best prices and prompt delivery. Please feel free to contact us at any time. We will reply you when we receive your inquiries. Please note that samples are available before we start our business.